PCB Piezotronics 443B02 Installation And Operating Manual page 51

Modular ac line power supply
Table of Contents

Advertisement

Command: TEDR
Purpose:
Response:
Command String:
Description:
Response String:
Example:
Manual 19854 Rev. D
ECN 50523
Returns data stored in the TEDS sensor.
Note: For sensors programmed with "Accelerometer, transfer function v0.91"
(UTID 116225), use MTED in conjunction with TEDR to retrieve all of the stored
data.
The module responds with a string of data delimited with semicolons.
XYCOZTEDR
X— Rack number of the target module (Range: ASCII printable characters, 0
through 3)
Y— Slot number of the target module. (Range: ASCII printable characters, 0
through 7)
C0Z— Module type for the target. Specifying "C01" in this argument addresses
a 443B101, specifying "C02" addresses a 443B102.
TEDR— Commands the module to return data from the TEDS sensor.
Manufacturer Model Number; Serial Number; Sensitivity; Reference Frequency;
Calibration Date; Low Frequency Response; Output Phase;
Sensitivity Direction; Measurement ID; User Data;
Command String:
"04C01TEDR"
Return String:
PCB 333M07; SN 17704; 100.2 mV/g; F ref 99.6; cal'd
3/21/2001; F hp 0.025 Hz; phase 0; sens dir N/A;
meas ID 0; test sample 4;
The 443B101 in rack 0 and slot 4 is commanded to return the data stored in the
TEDS sensor connected to its input.
443B101/443B102
31