Philips HF3329/60 User Manual page 24

Philips golite blu energy light hf3329
Hide thumbs Also See for HF3329/60:
Table of Contents

Advertisement

G
m
'
D
uiDance anD
anufacturer
S
eclaration
environment specified below. The user of the goLITE BLU should make sure it is used in such an environment.
i
T
iEC 60601 T
mmuniTy
EsT
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the goLITE BLU is used
exceeds the applicable RF compliance level above, the goLITE BLU should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the goLITE BLU.
b: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
- e
i
lectromaGnetic
mmunity
l
C
l
EsT
EvEl
omplianCE
EvEl
3 Vrms
3 V/m
: The goLITE BLU is intended for use in the electromagnetic
E
E
lECTromagnETiC
nvironmEnT
Portable and mobile RF communications equipment should be
used no closer to any part of the goLITE BLU, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey a , should be less than the
compliance level in each frequency range b .
Interference may occur in the vicinity of equipment marked with
the following symbol:
- g
uidanCE
23