Yamato DKG610 Instruction Manual page 50

Rapid heating / cooling oven
Table of Contents

Advertisement

5.2 Example of write communication
Example)
Request message: This message requests the Unit addressed to address03 to
set "SV setting to 135" (write 135).
Reply message from the Unit to this: Returns acknowledgement that the request
message was received.
☆Check that the message has been correctly written by separately reading out the data.
Code
① Start code
② Address
③ Request(Write)
④ Identifier
⑤ Numeric data
⑥ End code
⑦ BCC data Request
⑧ Acknowledgement code
Note 1) See "7.Identifier/command list" on page 49.
Note 2) For ASCII codes, see "8.ASCII code list" on page 51.
Write request message (Transmitted from a host PC)
S
T
0
3
W
S
V
X
Reply message (Returned from the Unit)
S
T
0
X
Symbol・data
note 1)
Reply
4. Operating procedures
Option (RS485 communication function)
1
0
0
1
3
A
E
B
3
C
T
C
K
X
C
⑧ ⑥ ⑦
STX
03
W
SV1
00135
30H 30H 31H 33H 35H
ETX
ACK
47
E
B
5
T
C
X
C
⑥ ⑦
ASCII code note2)
02H
30H 33H
57H
53H 56H 31H
03H
56H
04H
06H

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dkg610vDkg650Dkg650vDkg810Dkg810vDkg850 ... Show all

Table of Contents