MEASURED TEMP.
HEATER
ALARM
AUTO STOP
AUTO START
SET
TEMP.
FIXED TEMP.
PROGRAM
OVER TEMP .
PROTECTOR
RUN
ENTER
STOP
FIXED
SUB
TIMER
PROGRAM
TEMP.
MENU
Request
for
Check
operation
MEASURED TEMP.
HEATER
ALARM
AUTO STOP
AUTO START
SET
TEMP.
FIXED TEMP.
PROGRAM
OVER TEMP .
PROTECTOR
RUN
ENTER
STOP
FIXED
SUB
TIMER
PROGRAM
TEMP.
MENU
4. Operating Procedures
⑨ Press ENTER key. The character t-1
set time for the first step will be displayed and the current set
time flashes.
※Before setting a time, you need to know the temperature
increasing (or decreasing) capacity of the unit.
※As an example, about 45 minutes are required to increase
from room temperature to 260℃ for the model DKG610.
Thus as a rough guideline, temperature will rise by about
5℃ for one minute. In actual setting, set a time slightly
longer including a time necessary for stabilizing.
※Maximum settable time for the timer of each step is 999
hours 50 minutes.
⑩ When you have set a time, press
The temperature setting character Sv-2 for the second step is
displayed.
Then repeat the same steps as per the program sheet to input
temperatures and time for each step.
⑪ When you want to repeat a program pattern in the middle of a
program as in the example (program repeat), a special
procedures are necessary. In such a case, first set a time (t-7
in the example) for the step you want to perform repeat
operation (step 7 in the example) and press
without pressing
function setting mode.
※For operation and registration (input) procedures for the
Program
Repeat
operation" on P.31.
⑫ When you have set temperature and time for the final step,
the screen will return to the initial setting screen.
Be sure to perform no-load operation to check whether the set
temperature and the set time are correct before attempting actual
operation with specimens.
5.Starting Program operation
Press
RUN/STOP
operation you have set will start.
The Program operation lamp comes on and the Set Temperature
indicator will show steps being operated from the step St-1
first.
※You can press the ▼ key during operation to check the set
temperature and the remaining time for each step being operated on
the set temperature indicator.
29
ENTER
key. Now you can enter the Repeat
ENTER
function,
follow
key for about one second. The Program
Creating a program
that means the
key.
SUB MENU
key
"Program
Repeat

Advertisement