Temperature
increase/decrease
time for model DKG
Condition:Room temperature of 23℃
DKG610/650
Rise time
Fall time
260℃
20
250℃
30
200℃
25
20
150℃
100℃
15
50℃
10
Creating a program
Temp
260℃
200℃
100℃
0℃
Step
1
2
Temp. ℃
100
100
Time
0.15
0.30
MEASURED TEMP.
HEATER
ALARM
AUTO STOP
AUTO START
SET
TEMP.
FIXED TEMP.
PROGRAM
OVER TEMP .
PROTECTOR
RUN
ENTER
STOP
FIXED
SUB
TIMER
PROGRAM
MENU
TEMP.
The rough temperature increase/decrease time by 50℃ for model
DKG are as follows.
Numeric values mean required time between different temperatures.
〔Example: For DKG610, the time required to increase temperature
from 100℃ to 150℃ would be about 20 minutes.〕Be sure to set an
appropriate time after conducting a trial run since time required for
increasing/decreasing temperature will change depending on the
exhaust air amount and loads and a time for the temperature to
stabilize after the set temperature is attained shall be added or the
increase/decrease time may be needed to extend.
No-load Exhaust port full open
DKG810/850
Rise time
Fall time
-
20
-
20
30
10
35
30
35
40
25
35
45
20
50
80
10
105
Description here uses an example of a program pattern below.
1.Example program pattern
3
4
5
6
150
150
200
200
0.20
1.30
0.25
3.00
2. Turn power on(Turn ELB ON(|))
When you turn power on, the initial values will be displayed for
about four seconds and the screen will change to the initial setting
screen, and each of indicators displays the current in-bath
temperature, operation mode characters, and overheat prevention
set temperature.
3.Selecting a program mode and a program pattern
① Press
PROGRAM
The set temperature indicator displays the program mode use
in the last session.
Pressing
flash.
Pressing
mode flash.
4. Operating Procedures
DKG610/650
Rise time
25℃⇒
45
260℃
260 ℃
50℃
Repeat function used
7
No need to add the number of steps
50
3.00
key.
key again will make the program mode
PROGRAM
PROGRAM
key again will make the next program
27
Creating a program
〔Unit:minutes〕
DKG810/850
Fall time
Fall time
Rise
(Exhaust port
(Exhaust port
time
full open)
full open)
50
170
170
(30)
(40)
8
50
3.00