Specifications
Temperature output
(ANALOG)
Time up output
(TIME UP)
Temperature(℃)
0
20
40
60
80
100
120
140
160
180
200
220
240
260
270
※
<Reference value when voltage is input to the temperature recorder>
Converted voltage when shunt resistances (250Ω) are connected to the recorder input terminals
(V=RI:Ohm's law)
parallel
Unit:Temp. output terminal
(Current output)
4. Operating procedures
Outputs a voltage (DC) according to the measured temperature
Output temperature range:0~270℃
Output current :4~20mA
Output accuracy :±2℃
Connection
:M4 screw terminal block
Outputs on time-up of AUTO STOP, AUTO START or QUICK
AUTO STOP or at program end.
A-contact(relay contact)
Contact capacity : AC250V 3A(resistance load)
DC30V 3A(resistance load)
Connection
: M4 screw terminal block
Output current(mA)
4.00
5.19
6.37
7.56
8.74
9.93
11.11
12.30
13.48
14.67
15.85
17.04
18.22
19.41
20.00
38
Useful functions (Output terminal)
Converted voltage examples
1.00
1.30
1.59
1.89
2.19
2.48
2.78
3.08
3.37
3.67
3.96
4.26
4.56
4.85
5.00
Shunt resistance
Parallel onnection
Recorder input terminal
(Voltage input)
※(V)