Regular Source Checks; Cleaning Instructions - Thermo Scientific SAM12 Manual

Small articles monitor
Table of Contents

Advertisement

Issue: 1.0
Thermo Fisher Scientific
Should the results be outside these limits, the SAM12
operational parameters require revision (see
Procedure (page
9-1)). Calibration procedures are described in
Calibration Procedure (page
Carry out a regular source check after a 100 sec background
update.

Regular Source checks

Daily or weekly source checks are advisable, using a
Calibration check as described in
The check source is placed in the calibration position, usually 3
inches from the bottom of the measuring volume.
The alarm response should be 2 out of 3 or greater than the
probability of detection for a "one alarm level" source.

Cleaning Instructions

WARNING: Ensure the mains supply is isolated before
cleaning.
The display can be wiped clean with a dry cloth. Smudges may
require the application of the Screen cleaner provided when the
instrument was delivered.
The cubical should be cleaned using a mild detergent.
Make sure the equipment is completely dry before re-
connecting the supply.
Setting Up
10-1).
Cal Check (page
SAM12 User Manual
Routine Checks
5-19).