Operating Instructions
5-58
SAM12 User Manual
information regarding Lock Front
Door on Alarm.
If the High Level Alarm - see
Alarms (page 5-12)
- is exceeded,
the "High Level Alarm" message
will display.
Note that this message is user-
definable, see
Messages (page 5-
15).
If the Lock Front Door on Alarm
is set to High or All Alarms, the
door will be locked and can only
be opened with intervention by
an Administrator.
See
Doors (page 5-13)
information regarding Lock Front
Door on Alarm.
On a two-door SAM12, the front
door will be locked and back
door will be unlocked.
Also, a high alarm will sound for
5 seconds.
If the count rate from any
detector exceeds 95% of the
Amplifier "dead-time" an "Over
Range" message is displayed
stating that the measurement was
over the operational range of the
instrument.
If the CCM board is installed and
the CCM option - see
(page 5-6)
- is selected, when the
CCM alarm – see
Alarms (page
5-12)
- is exceeded, the "Co60
ALARM" message is displayed.
Note that this message is user-
definable, see
Messages (page 5-
15)
for further information.
Also, an alarm will sound for 3
seconds even if the door is
opened.
If the Lock Front Door on Alarm
is set to All Alarms, the door will
be locked and can only be opened
for more
Options
Issue: 1.0