Issue: 1.0
Doors
Thermo Fisher Scientific
Default: 5000.0 dpm
High Level Alarm
This is the detected activity at which SAM12 will
(default unit)
indicate an High Level Alarm
The Enabled tickbox must be checked for this
option to take effect.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
CCM Alarm (default
This field is enabled if CCM Enabled is ticked - see
unit)
Options (page
This is the detected activity at which SAM12 will
indicate a CCM Alarm.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
CCM High Level
This field is enabled if CCM Enabled is ticked - see
Alarm (default unit)
Options (page
This is the detected activity at which SAM12 will
indicate a CCM High Level Alarm.
The Enabled tickbox must be checked for this
option to take effect.
Values: 0 to 10,000,000 (Any unit).
Default: Must be input
Select Set-Up|Doors to select the
Number of Doors, whether Locks
are fitted and their behaviour
Note that this section is set by at production and should only be
changed in consultation with Thermo Fisher Scientific.
Description of options:
Field Name
Description
Number of Doors
Use the Radio buttons to select either One Door or
Two Doors.
Default: Two
Locks Fitted :
Front Door
Tick this box if the Front Door has a lock fitted.
Also referred to as the "Dirty" door.
Default: ON
Rear Door
Note that this option is only visible if Two Doors is
selected in Number of Doors.
Tick this box if the Rear Door has a lock fitted.
Also referred to as the "Clean" door
Operating Instructions
5-6).
5-6).
SAM12 User Manual
5-13