Thermo Scientific SAM12 Manual page 88

Small articles monitor
Table of Contents

Advertisement

Operating Instructions
Out of Calibration
Establishing a new
Background
5-48
SAM12 User Manual
If the Calibration Required
Interval option has been
exceeded, then the Out of
Calibration message is displayed
– see
Params 1 (page
check the Calibration of the unit,
see
Cal Check (page
5-19).
Once the Self Checks have
passed, SAM12 will to acquire a
new background and the
"Updating Background" message
is displayed.
See
Notes regarding Changing
Background (page 5-51)
information regarding acquiring a
background.
When a new Background has
been established, the "Ready –
OPEN DOOR TO MONITOR
ARTICLES" message is
displayed. The SAM12 is now
ready to start monitoring articles.
If the front door is opened when
the SAM12 is establishing a new
background, background
counting stops, the "Close Door
To Resume" message is
displayed and an audible alarm
will sound. To continue with the
process, the door(s) must be shut.
If the back door is opened when
the SAM12 is establishing a new
background, background
counting stops, the "ILLEGAL
DOOR SEQUENCE" message is
displayed and an audible alarm
will sound. To continue with the
process, the door must be shut.
See
Doors (page 5-13)
information.
5-8). To
for more
for more
Issue: 1.0
Thermo Fisher Scientific