Operating Instructions
5-20
SAM12 User Manual
Select Calibration|Cal Check to
edit the Calibration setting
against a selected source and to
conduct a calibration check.
The Decay Corrected Activity is
displayed on this screen.
Note that the Calibrate button is not available to the Technician
role.
To edit Calibration Source
properties, use the direction
arrows to select a Calibration
Source. Once selected, press Edit
button to display the Edit Source
properties window:
Description of options:
Field Name
Description
Nuclide
This option is only available when creating a new
Calibration Source (see below).
This changes the displayed activity units for the
SAM12.
Source Name
Enter a Source name.
Last Calibration Date
Select a Calibration Date from the date picker.
Activity
Enter an Activity value.
Use the pre-defined window to select a unit from
one of the following:
Comment
Enter a Comment to describe the Source, if
required.
• dpm
• Bq
• kBq
• pCi
• nCi
• μCi
• mCi
• Ci