Technical Characteristics - Hologic Acessa ProVu User Manual

Table of Contents

Advertisement

Acessa
ProVu System User's Guide
®
18.3.

Technical Characteristics

Coagulation Mode Output at Various Settings into Various Loads
60.0
50.0
40.0
30.0
20.0
10.0
Approximate peak coag voltage vs. load resistance at all levels ("coag level" per the user interface).
Note: The load for coag will increase as the affected tissue cauterizes and dries. As the voltage level increases,
the effect of coag is more immediate. However, slower coagulation (lower levels) may be more effective.
Watts of Power vs. R
250.0
200.0
150.0
100.0
50.0
Note: The unit is limited at low load impedances by maximum available current and is limited at high load
impedances by maximum available voltage. 200 watts is only available over a load range of 50 Ω to 80 Ω. The
system will not ablate any loads over the system ablate high resistance alert of >511Ω.
CO 20-131 Revision: C
Coagulation Mode
0.0
0
500
(Ablate Modes)
load
Ablation Modes
0.0
0
500
1000
Resistance
1000
1500
Resistance
1500
2000
2500
PL-01-0040
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2000
2
1
Constant Power
Constant Temp