Dc/Mv Dc Adjustment Procedure; Dc Adjustment Procedure - Fluke 393 Calibration Manual

Trms clamp meter
Hide thumbs Also See for 393:
Table of Contents

Advertisement

393/393 FC
Calibration Manual
V DC/mV DC Adjustment Procedure
See
Table 5
for both the V dc/mV dc adjustment steps. When the Product shows SAUE on
the LCD, calibration is complete.
Step
Display
1.
C-08
2.
C-09
3.
C-10
4.
C-11
5.
C-08
6.
Save
V DC Adjustment Procedure
See
Table 6
for only the V dc adjustment steps. When the Product shows SAUE on the
LCD, calibration is complete.
Step
Display
1.
C-08
2.
C-09
3.
C-10
4.
C-11
5.
C-08
6.
Save
18
Table 5. V DC/mV DC Adjustment
Calibrator Output
0 V
600 V
0 V
0.5 V
STBY
---
Table 6. V DC Adjustment
Calibrator Output
0 V
600 V
STBY
STBY
STBY
---
Operation
Push
Push
Push
Push
Push e
---
Operation
Push
Push
Push I
Push I
Push e
---