374 FC/375 FC/376 FC and 902 FC
Calibration Manual
9.
For each calibration adjustment:
a.
Turn the rotary knob to select the function.
b.
Apply the required output from the source to the Product.
c.
Wait until each applied output stabilizes.
On the 37x FC Products, push to confirm the value and move to the next step in the Adjustment
Procedure. When you have completed all the steps, push
On the 902 FC Products, push
Procedure. When you have completed all the steps, push
10. To view the target calibration point, push .
11. When calibration is complete:
a.
Remove the power supply.
b.
Replace the batteries.
c.
Replace the battery compartment door and tighten the battery compartment screw.
VAC Adjustment Procedure
Step
LCD Display
1
C-00
2
C-01
3
C-02
4
C-03
5
C-04
6
C-05
7
C-06
8
C-07
9
Save
12
Figure 5. Current Calibration Setup
to confirm the value and move to the next step in the Adjustment
374 FC
375 FC
376 FC
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
to save the data to NVRAM storage.
to save the data to NVRAM storage.
902 FC
Calibrator Output
√
0 V, 0 Hz
√
600 V, 50 Hz
√
300 V, 50 Hz
√
300 V, 100 Hz
√
300 V, 200 Hz
√
300 V, 300 Hz
√
300 V, 400 Hz
√
300 V, 500 Hz
√
STBY
5522A
CALIBR ATOR
ghn50.eps