Vac Adjustment Procedure - Fluke 374 FC Calibration Manual

Clamp meter
Hide thumbs Also See for 374 FC:
Table of Contents

Advertisement

374 FC/375 FC/376 FC and 902 FC
Calibration Manual
9.
For each calibration adjustment:
a.
Turn the rotary knob to select the function.
b.
Apply the required output from the source to the Product.
c.
Wait until each applied output stabilizes.
On the 37x FC Products, push  to confirm the value and move to the next step in the Adjustment
Procedure. When you have completed all the steps, push
On the 902 FC Products, push
Procedure. When you have completed all the steps, push
10. To view the target calibration point, push .
11. When calibration is complete:
a.
Remove the power supply.
b.
Replace the batteries.
c.
Replace the battery compartment door and tighten the battery compartment screw.

VAC Adjustment Procedure

Step
LCD Display
1
C-00
2
C-01
3
C-02
4
C-03
5
C-04
6
C-05
7
C-06
8
C-07
9
Save
12
Figure 5. Current Calibration Setup
to confirm the value and move to the next step in the Adjustment

374 FC
375 FC
376 FC
to save the data to NVRAM storage.
to save the data to NVRAM storage.
902 FC
Calibrator Output
0 V, 0 Hz
600 V, 50 Hz
300 V, 50 Hz
300 V, 100 Hz
300 V, 200 Hz
300 V, 300 Hz
300 V, 400 Hz
300 V, 500 Hz
STBY
5522A
CALIBR ATOR
ghn50.eps