Chapter 2: Aperio GT 450 DX Network Architecture
Data Type
Image Data
The Scanner sends DICOM image data to the DICOM Converter. The data is sent using
TLS encryption.
Configure the communication between the scanner and the DICOM converter using the
Hostname and Port settings on the Images configuration page.
The DICOM Converter sends the image data (either as a converted SVS file, or as raw
DICOM data) to the image and data management system (IDMS) on the DSR Server. The
data is sent using SMB3 Encryption.
Configure the communication between the DICOM converter and the DSR using the File
Location setting on the Images page.
Images can be sent to viewing stations connected to the DSR.
Scanner
The scanner sends a call to the SAM DX DataServer to request configuration data. The
Configuration
SAM DX DataServer returns the configuration data to the scanner. The data is sent using
Data
TLS Encryption. Communication between the scanner and the SAM DX DataServer is
configured on the scanner.
The SAM DX DataServer stores the configuration data on the SQL Server Database on the
SAM DX Server.
The SAM DX DataServer displays the configuration data through the SAM DX web page.
Time
Timeclock synchronization between SAM DX and Multiple Scanners is maintained using
Synchronization
network time protocol.
Image Metadata
The Scanner sends Image Metadata to the SAM DX DataServer. The data is sent using
TLS encryption. Communication between the scanner and the SAM DX DataServer is
configured on the scanner.
The SAM DX DataServer sends image metadata to the IDMS located on the DSR. The
data is sent using TLS encryption.
Configure the communication between the SAM DX DataServer and the scanner using
the Hostname and Port settings on the DSR page.
Messaging and
The scanner sends logs and event data to the Mirth Connect Server. No sensitive data is
Event Data
transferred.
Configure the communication between the scanner and the Mirth Connect Server on the
Event Handling configuration page.
The Mirth Connect Server copies critical event and error data to the SAM DX DataServer,
and then the SAM DX DataServer sends this data to the SQL database. This is the data
reported out via the SAM DX Event Logs.
The SAM DX DataServer displays the event data through the SAM DX web page.
The Mirth Connect Server processes the Log data and appends the Event Log, which
resides on the file system. The communication between Mirth and the Event Log is
configured within the Mirth Application setup. It is not accessible through SAM DX.
"Scanner Configuration Settings" on page 23 provides information on how to configure the various connections between the
components and services through the SAM DX interface.
18
Aperio GT 450 DX IT Manager and Lab Administrator Guide, Revision B © Leica Biosystems Imaging, Inc. 2022
Description
Port
TCP 2762
UDP 137,
138
TCP 139,
445
80, 443
44386
UDP 123
44386
6662, 6663

Advertisement