Olympus DP74 Instructions Manual page 54

Microscope digital camera
Table of Contents

Advertisement

Proper selection of the power supply cord
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to
"Specifications" and "Certified Cord" below:
Caution : In case you use a non-approved power supply cord for Olympus products, Olympus can no longer
Voltage rating
Current rating
Temperature rating
Length
Fittings configuration
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with
an agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of the agencies
listed in Table 1. In case you are unable to buy locally the power supply cord which is approved by one of the agencies
mentioned in Table 1, please use replacements approved by any other equivalent and authorized agencies in your
country.
50
warrant the electrical safety of the equipment.
125 V AC (for 100-120 V AC area) or, 250 V AC (for 220-240 V AC area)
6 A minimum
60 ˚C minimum
3.05 m maximum
Grounding type attachment plug cap. Opposite terminates in molded-on IEC
configuration appliance coupling.
Specifications
Table 1 Certified cord
DP74