Field number 22
22
} If the acquired field is as shown in left figure, use a camera adapter
0.63X (FN 14)
0.5X (FN 17.6)
1X (FN 8.8)
} The view field range illustrated in left figure is a guide. Check the
CAUTION
c
having magnification of 0.5X to 1X. (If a 0.35X camera adapter is
used, the peripheral part of the image will be obscured.)
If a C-mount camera adapter from other manufacturer than Olym-
pus is used, the optical performance of the system may not be
manifested fully.
actual view field range by observing the specimen equipped with a
scaling function, etc. before use.
• Be careful in using other manufacturer's C-mount camera
adapter or C-mount lens having a thread length
mm. Otherwise, the threaded section will hit the inside of the
camera head and cause damage to it.
• Be sure to adjust the parfocality between the C-mount camera
adapter and the eyepieces. Otherwise, the image through the
camera and the image through the eyepieces are not parfocal.
For the parfocality adjustment method, refer to the instruction
manual for the camera adapter in use.