Managing network connections in client mode
configuration
The DR detector is configured for client mode.
Figure 24: Client mode
Each workstation has an access point or a System Control Unit.
Multiple DR detectors can be connected to a MUSICA Acquisition Workstation
simultaneously.
The DR detector can be used for examinations on multiple MUSICA
Acquisition Workstation.
Connecting to another MUSICA Acquisition Workstation
(client mode)
The DR detector is configured in client mode.
To connect the DR detector to another MUSICA Acquisition Workstation:
1. Take the DR detector near the MUSICA Acquisition Workstation.
2. Press and hold the AP mode button for 3 seconds.
Figure 25: AP mode button
The detector status display shows "AP scan", indicating that the wireless
network is being scanned. From a list of access points that is configured
upon installation, the one with the highest signal strength is automatically
selected. After a while the detector display shows "Change AP".
Note that if the DR detector was already connected, that access point will
be ignored, even if it has the highest signal strength. To avoid confusion,
always check the SSID name of the access point to which the DR detector
is connected.
3. Wait until the connection to the MUSICA Acquisition Workstation is
established.
The detector status display shows the signal strength.