AGFA XD 17 User Manual page 77

Hide thumbs Also See for XD 17:
Table of Contents

Advertisement

Figure 16: Power button
During startup the power indicator is blinking green. After startup the
power indicator is green.
6. Check the DR detector status icon on the DR Detector Switch.
If the displayed status is error, following step is required:
• If the DR detector is configured in client mode and the displayed status
is error, connect the DR detector to the MUSICA Acquisition
Workstation.
• If the DR detector is configured in access point mode, use the Windows
wifi setting to connect the MUSICA Acquisition Workstation to the
wireless network of the DR detector.
The DR detector is ready.
Before exposure make sure to check the equipment daily and confirm that it
works properly.
Related Links
Viewing the detector status
Managing network connections in client mode configuration
Managing network connections in access point mode configuration
DR detector not ready for exposure
on page 97
on page 115
XD 17, XD⁺17 | Getting started | 77
on page 106
on page 108