114 | XD 17, XD⁺17 | Problem solving

Artifact in DR Detector images

Details
Cause
Brief Solution
0434A EN 20210426 1355
An artifact is visible in the images produced by a DR
Detector.
Exposure conditions have changed significantly since
latest calibration.
Perform calibration of the DR Detector.
For details, refer to the DR Detector Calibration Key
User Manual (document 0134).