Summary of Contents for Shodex Asahipak NH2P-40 2E
Page 1
Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex Asahipak NH2P series is polymer-based chromatography columns modified with amino group. The column is suitable for the analyses that require HILIC or ion exchange mode separations.
Page 2
Range Range (°C) Recommended Maximum (Per Column) Asahipak NH2P-40 2E 0.1 - 0.2 2 - 13 15 - 50 Usable solvents are listed below. (1) A mixture of water and acetonitrile or water and ethanol of any ratio can be used. Up to 100 % water, acetonitrile, or ethanol can be used.
Page 3
This may result in creating an empty space in the column and deteriorates the column. Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
· Strong alkaline solvents such as 0.1-M aqueous sodium hydroxide solution can damage the detector cell. Do not connect the detector while regenerating the column and collect elute directly from the column outlet to the waste container. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.
Need help?
Do you have a question about the Asahipak NH2P-40 2E and is the answer not in the manual?
Questions and answers