Summary of Contents for Shodex Asahipak NH2P-50 2D
Page 1
Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex Asahipak NH2P series is a polymer-based chromatography column modified with amino group. The column is suitable for the analyses that require HILIC or ion exchange mode separations.
Page 2
Pressure (MPa) pH Range Range (°C) Recommended Maximum (Per Column) Asahipak NH2P-50 2D 0.1 - 0.2 2 - 13 4 - 50 Asahipak NH2P-50G 2A Usable solvents are listed below. (1) A mixture of water and acetonitrile or water and ethanol of any ratio can be used. Up to 100 % water, acetonitrile, or ethanol can be used.
Page 3
This may result in creating an empty space in the column and deteriorates the column. Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
Method 2: Adsorption of metal ions or in case Method 1 did not work Washing Solution Cleaning Time 6 min 0.1 M HNO aq. solution 75 min 6 min 0.1 M NaOH aq. solution 75 min 12 min Eluent 40 min SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.