Connect Lid Strap (Model 7406/7407); Connect To Electrical Supply - Thermo Scientific CRYO Plus 1 Operating And Maintenance Manual

Liquid nitrogen storage system
Table of Contents

Advertisement

Section 1
Quick Start-Up
Connect Lid Strap
(Model 7406/7407)
Connect to Electrical
1-4
7007400 Cryoplus
Included with each unit is a lid strap for the user's convenience.
1. Remove the protective white nylon screws from the areas indicated.
Discard these screws.
2. Install the strap as shown below, using the screws included with the
strap.
Figure 1-3. Lid Strap Installation
With the power switch turned OFF, connect the unit to a grounded
electrical outlet. See the data plate on the back of the unit, or the electrical
Supply
schematics included in this manual, for voltage and full load amps.
The power switch on the back of the unit is the mains disconnect and is
also a reset-type circuit breaker. If an overload condition occurs, the built-
in circuit breaker will trip and the power switch will turn off. Turning the
power switch on resets the circuit breaker. If the circuit breaker trips again
within a short time period, the unit should be checked by a qualified
electrician.
Warning Use only a grounded electrical receptacle. Failure to ground the
unit can result in serious injury.