condition that triggers the capturing when the force gauge is in curve
capture mode. When the maximum number of data has been
recorded or capturing was discontinued early, a capture report is
created and saved. The curve is deleted when you leave capture
mode.
The baud rate for the serial interface can be set to a value between
4800 and 230400 bps.
This setting will only be effective after restarting the force gauge.
Note: To make sure that all data are retrieved when the device is
connected to a computer, the baud rate should be set as follows:
12 Hz: ≥9600 bps
25 Hz: ≥14400 bps
Baud Rate
50 Hz: ≥19200 bps
100 Hz: ≥28800 bps
200 Hz: ≥38400 bps
400 Hz: ≥57600 bps
800 Hz: ≥115200 bps
1600 Hz: ≥230400 bps
Due to the limited speed of serial interfaces, some data get lost when
transferred to a PC if the sampling rate is higher than 800 Hz. The
readings will, however, not get lost in the device.
Here, you can set the display angle. You can select 0 or 180 °.
Display Angle
This function also reduces the energy consumption. The brightness of
Auto Backlight
the display backlight will automatically be turned off when the meter
has not been used for a certain period of time.
Now Voltage
This window shows the current battery level.
You can reset the device to factory default settings, e. g. if you have
Reset
made an incorrect setting or if other problems with the settings
occurred.
In this window, you can change the output language. English or
Language
German can be selected.
In this window, you can see the circuit diagram of the input port (see
Connection
chapter 3.2).