..........2 B. Soak the Electrode ...2 C. Setup and Calibrate the Electrode and Meter PCE-ORP 3 OVERVIEW..........3 ..........3 A. Meter Description PCE-ORP 3 Graphical Display .......3 OPERATION MODES AND KEYPAD OPERATIONS..4 ...........4 A. Operation Modes ..........4 B. Keypad Operations USING PCE-ORP 3............5...
2. Soak the ORP electrode in standard solution or distilled water for 10 minutes before first use or after storage. C. Setup and Calibrate the Electrode and Meter PCE-ORP 3 must be setup and calibrated before your first use. Please follow the instructions detailed in section USING VPCE-...
OPERATION MODES AND KEYPAD OPERATIONS A. Operation Modes . PCE-ORP 3 meter has 7 operation modes: 1. Measure Mode.--Measure Mode is used to make all absolute or relative ORP and temperature measurements. 2. Calibration Mode.--Calibration Mode is used to perform offset voltage adjust.
Erase 0 second Exits Erase Mode USING PCE-ORP 3 A. Power ON/OFF & Shutdown Press “HOLD” key for 2 seconds or more. The unit will turn on and enter the “Measure Mode”. Repeat the process to turn off the unit.
XX. The unit will automatically returns to “Measure Mode” and ready to perform other task. Note: The PCE-ORP 3 is equipped with a non-volatile memory that can store up to 50 different sets of readings...
(ABSolute & RELative mV, temperature, Time and date).Non- volatile memory will be retained even if power is lost. When all 50 memory locations are used up, the next set of data saved will overwrite the data stored in the first location and so on. G.
ERROR DISPLAYS AND TROUBLESHOOTING DISPLAY Possible cause(s) Display Mode [Action(s)] Display "OVER" Measure a. Temperature outside –10.0 to + Value 100.0°C range. “UNDER” [Bring solution to a lower/higher temperature.] b. Bad sensor. [Replace new electrode module .] "OVER" -10.0~ Measure ORP value outside -1100 to +1100 100.0°C mV value.