You can set a value between 1 and 1280 seconds. This value
represents the duration of curve capturing in capture mode which
depends on the sampling rate:
Sampling rate 6 Hz: 1 ~ 1280 seconds
Sampling rate 12 Hz: 1 ~ 640 seconds
Capture Length
Sampling rate 25 Hz: 1 ~ 320 seconds
Sampling rate 50 Hz: 1 ~ 160 seconds
Sampling rate 100 Hz: 1 ~ 80 seconds
Sampling rate 200 Hz: 1 ~ 40 seconds
Sampling rate 400 Hz: 1 ~ 20 seconds
Sampling rate 800 Hz: 1 ~ 10 seconds
Sampling rate 1600 Hz: 1 ~ 5 seconds
You can set a value between 1 and 99999. This setting is used during
peak value measurement to determine the first peak value. When you
press the Arrow Right key, a new peak value measurement will start.
Meanwhile, the values peak-to-peak (Vmax), valley-to-peak (Vmin)
F/P Boundary
and new peak (Vnew) are updated continuously. For example, if 10 is
set as the criterion, Vmax or Vmin will be counted as the first peak
value when the absolute value of (Vmax - Vnew) or (Vmin - Vnew) is
above 10.
This port is used to control the real-time data transfer. The following
parameters can be set:
Prohibit: The real-time data transfer of the serial interface is
prohibited.
Key/Order: A single output will take place when you press the Up key
or when an output command is received. When the force gauge is
Serial Port
connected to a computer, the programmes on the computer will
automatically disable the output function.
Change: A single output will take place when the measuring data
change.
Stabilize: A single output will take place when the reading stabilises.
Consecutive: The measuring data are transferred without interruption.
This function reduces the energy consumption. The force gauge will
Auto Power Off
automatically power off when it has not been used for a certain period
of time.
Max Charge V
This window shows the maximum voltage of the battery.
Here, you can delete saved measurement reports and curves.
Clear Storage
Important information:
When the memory is full, all data will automatically be deleted to
enable new data to be saved.
Factory Test
Only relevant for customer service.
This window shows the serial number of the device which cannot be
S/N
changed.
Here, you can set a value between -99999 and +99999. The range of
Capture Trigger
values depends on the set unit. This parameter stipulates the