AGFA DX-D 10C User Manual page 39

Hide thumbs Also See for DX-D 10C:
Table of Contents

Advertisement

Maximum load
Maximum total load
Maximum Linear Dose
using RQA5
Minimum Modulation Transfer Function (MTF) using RQA5
1 lp/mm
2 lp/mm
3 lp/mm
Minimum Detective Quantum Efficiency (DQE) using RQA5 at 2.1 µGy
dose level with reduced cycle time
0.5 lp/mm
1 lp/mm
2 lp/mm
3 lp/mm
Minimum Signal Noise Ratio for 1mR
SNR
Conversion screen
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G | Technical Data | 39
100 kg
DX-D 10G
DX-D 20G
75 µGy
0.45
0.15
0.05
0.23
0.18
0.08
0.02
115:1
GOS
DX-D 10C
DX-D 20C
50 µGy
0.50
0.25
0.12
0.45
0.35
0.20
0.10
120:1
CsI
0129E EN 20180312 0812

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dx-d 10gDx-d 20cDx-d 20g

Table of Contents