24 | DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G | Safety Directions
0129E EN 20180312 0812
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
CAUTION:
Position the DR Detector control unit so that it is possible to
disconnect the mains power connection if required.
CAUTION:
When rolling or pulling the cable of the DR Detector, take care
not to wind the cable too tightly. The bending radius of the cable
should not be smaller than 10 cm
CAUTION:
Observe great care when handling the DR Detector. The detector
is shock sensitive and drops should be avoided. Warranty will be
void if it is obvious that operating conditions are not met.
If the DR Detector has been dropped:
1. Visually check the DR Detector for deformations.
2. Perform a calibration of the DR Detector. For instructions, refer
to the DX-D DR Detector Calibration Key User Manual
(document 0134).
3. Perform a flat field exposure and check the image for visible
artifacts. Typical flat field exposure settings are 75 kV, 10 µGy,
large focus and using 1.5 mm Cu filter without grid.
CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.
CAUTION:
If weight is put on the DR Detector, it must lie on a flat surface.
CAUTION:
The load on the DR Detector must not exceed 100 kg.
CAUTION:
Excessive ambient temperature may impact the performance
and cause permanent damage to the device. If ambient
temperature and humidity is outside the range of 10 - 35 °C and
30 - 75% RH, do not operate the system or use air conditioning.