Load cells with four-wire system
Do not shorten or length connecting cables in the four-wire system, because the cable resistance
is temperature compensated. When the length of the connecting cable is altered, the input and
output resistances change. This change can be corrected by adjusting the scale, but
temperature-dependent resistance changes are not compensated for the missing or additional
length of cable.
Figure 5-1
Connecting principle for load cells with four-wire system
Note
No calibration approval
In scales requiring official calibration, the connecting cables for load cells in a four-wire system
must not be shortened or lengthened.
SIWAREX WL200
Operating Instructions, 12/2021, A5E02199611B-13
Connecting
5.2 Connecting principle
27