Verification; Nibp Verification; Battery Operation Verification; Co 2 Operation Verification - Mindray Spectrum Service Manual

Datascope
Table of Contents

Advertisement

Calibration Procedure
4.6.5
SpO
1. Set the display waveform 4 as Pleth. Set the HR source to Auto.
2. Verify that the SpO
3. Connect the SpO
4. Apply sensor to finger.
5. Verify SpO
4.6.6

NIBP Verification

1. Connect the Adult cuff connector to the NIBP hose. Attach the NIBP hose to the Cuff
2. Apply cuff and press the START key.
3. Verify the pump motor starts to pump and inflates the cuff to 180 mmhg (Adult).
4. Verify the reading on screen.
4.6.7

Battery Operation Verification

1. If batteries are installed in the unit remove them.
2. Verify the unit functions on Line power correctly.
3. Install the two batteries in the appropriate slots located on the left side of the monitor.
4. Remove the line cord from the unit. Verify the unit operation is not interrupted.
5. Remove one of the batteries and verify the unit still operates. Verify the second battery
4.6.8
CO
1. Connect the FilterLine
2. Attach a can of Calibration Gas (P/N 0075-00-0033-01) to the Filterline Short Term
4.6.9

Cardiac Output Verification

Setup
1. Access the Installation Menu by pressing and holding the DISCHARGE key
2. Select Enable EPM and change to Yes.
3. Select the weight units to kgs.
4. Select the height units to cms.
5. Save the current settings and turn unit OFF.
®
Spectrum
/Spectrum OR
Service Manual

Verification

2
message is displaying SpO
2
sensor to the panel connector. Verify the SpO
2
Sensor Off or SpO
Initializing.
2
displays the pleth waveform, and the SpO
2
Verify the HR source is SpO
connection the left side on the monitor.
The cuff will begin to deflate and obtain a blood pressure reading of Sys/ Dia/ Mean in
about 20 to 30 seconds after peak pressure is obtained.
operates, if installed alone.
Operation Verification
2
®
Short Term assembly to the input port of the CO
the left side of the monitor.
assembly. Feed gas into monitor and verify the ETCO
readings are displayed on the screen.
®
(Spectrum
) or the TRENDS key (Spectrum OR
0070-10-0556-02
No Sensor in Message area "B".
2
and a beep tone is present.
2
, Inspired CO
2
) while powering ON the monitor.
message changes to
2
indicates a valid reading.
2
connector on
2
and respiration
2
Verification
4 - 23