Cleaning and Disinfecting, continued
Hardware
CAUTION: Before cleaning or disinfecting, remove the cover and the support pad from the hardware.
Cleaning: Hand wash in warm (40°C/104°F) water with soap and a sponge. Rinse with clean water
and wipe dry with a clean cloth. Dry completely before use.
Disinfecting: Wipe gently with a cloth dampened with a household disinfectant. Rinse with clean,
warm water. Wipe dry with a clean cloth. Dry completely before use.
ROHO Air Insert
WARNINGS:
- Cleaning and disinfecting are separate processes. Cleaning must precede disinfection. If the product becomes
soiled, clean, disinfect, and check the product for proper functioning.
- DO NOT allow water or cleaning solution to enter the air insert. Make sure the inflation valve is closed.
- Remove the air insert from the foam base before cleaning or disinfecting.
CAUTIONS: Use of the following may cause damage to the air insert: abrasives (steel wool, scouring pads);
caustic, automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents,
including acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover, oil-based
lotions, lanolin; ozone gas; ultraviolet light cleaning methods. DO NOT expose to sunlight.
Cleaning: Remove the air insert from the foam base. Close the inflation valve and place in a large
sink. Hand wash, using mild liquid hand soap, hand dish-washing liquid, laundry detergent, or a
multipurpose detergent (follow product label instructions). Gently scrub all surfaces, using a soft
plastic bristle brush, a sponge or a cloth. Rinse thoroughly with clean water. Let dry completely before
use.
Disinfecting: Follow the hand wash instructions, using 1 part household liquid bleach per 9
parts water. Keep the product wet with the bleach solution for 10 minutes. Rinse thoroughly
with clean water. Let dry completely before use.
Care Symbols
Wipe with a damp
cloth.
Hand wash warm.
Machine wash
40C
warm, gentle.

Maintenance

WARNING: DO NOT continue to use the seating system if any part is broken, cannot be tightened or is not
functioning properly. Periodically, the seating system should be thoroughly inspected by a qualified professional
or healthcare provider.
Weekly, inspect all components of the equipment.
Check and tighten all fasteners and adjustment points on the mounting hardware.
Inspect all materials for wear, such as stretching, fraying, fractures, and bends.
Components that show excessive wear should be replaced. For information on replacements, refer to the
"Warranty" section in this manual.
Tru-Shape Operation Manual
Hand wash.
Tumble dry low.
Machine wash hot, normal, at
60C
the temperature shown.
Dry flat.
Bleach (1 part bleach: 8
parts water).
Bleach (1 part bleach: 9
parts water).
9
Line dry.
Dry flat in the shade.