3.2.7 Prepare the Tabletop for Moving (A, W,Wi)

WARNING:
Do not tilt the unit to the vertical position until the Table End Bracket is properly
installed (step 4 below).
Before the Discovery can be tilted to vertical, you must slide the table to the left and clamp
it so it clears the floor when the unit is tilted to vertical. To accomplish this, do the
following:
1. Remove the tabletop pad.
2. Remove the three Phillips screws holding the tabletop. Two of the screws are at
the left corners and one at the right back corner looking down on the tabletop.
3. Slide the tabletop to the right until it clears the table base panel.
4. Referring to Figure 3-10, adjust the Table End Stop as pictured.
Figure 3-10. Table End Stop (Models A, W, and Wi

3.2.8 Move Unit To Destination

Follow the procedure below to move the unit:
1. Remove the four shipping brackets (Figure 3-11) as follows: loosen the screws
that hold the auxiliary caster bars (located under each bracket), remove the lag
bolts and remove the bracket. Then, re-tighten the screws that hold the auxiliary
caster bar to the frame.
Bracket
3-12
Discovery QDR Series Technical Manual
Figure 3-11. Shipping Bracket