Section 6 Cleaning & Sterilization
30
• After cleaning/sterilizing the Y-Sensor, verify the sensor is physically intact, with no
broken/frayed wires or damaged parts. Ensure the connectors are clean and dry, with no
signs of contamination or corrosion. Do not use a broken or damaged sensor or one with
wet, contaminated, or corroded connectors.
• Perform a Y-Sensor Quick Check test before returning the sensor to patient use. Refer to
"Y-Sensor Quick Check" on page 18.

Y-Strip Tapes

• Treat Y-Strip Tapes in accordance with hospital protocol for single-patient use.

Ear Clip

• Clean the ear clip with a cloth dampened with 70% isoprophyl alcohol. After cleaning
wipe the ear clip down thoroughly with a clean water dampened cloth to rinse.
• Do not immerse the ear clip.
01-May-96
Model 511
User's Manual