Model 511
Model 511 User's Manual
Single Patient Use SpO
Intended Use:
The Single Patient Use SpO
monitoring adult, pediatric or neonatal patients with Novametrix Pulse Oximeters
(SuperBright™ series). These sensors are used with the DB-9 connector cable.
Use the Single Patient Use sensor and DB-9 connector cable only with
WARNING:
Novametrix SuperBright compatible pulse oximeters. Use with any other device may result
in equipment damage or patient injury.
These SpO
CAUTION:
reapplied to various sites on the same patient but should not be used on multiple patients.
Do not attempt to clean or disinfect the sensor, as system performance will be
compromised.
The Single Patient Use sensor should be discarded if sensor integrity becomes a
NOTE:
question.
Instructions for Use:
Connecting the sensor:
1. Attach the DB-9 to OxySnap connector cable to the Model 511. Align the arrows on
the OxySnap connectors and press the connectors together.
DB-9
Connector
Cable
Sensors
2
sensors (Catalog Nos. 6480 and 6455) can be used when
2
sensors are intended for single patient use. The sensors can be
2
Model 511
01-May-96
Single Patient Use SpO
Sensors
2
19