10:1 Divider

The 10:1 divider fits over the tip of the probe and provides the following changes to the probe's
operating parameters:
• Increases (by a factor of 10) the input voltage at which 1 dB compression occurs.
• Decreases the input capacitance without changing input resistance, thereby decreasing capacitive
loading.
Two or more dividers may be cascaded to provide higher divide ratios.

Operating Characteristics

Divider Ratio
Input Capacitance
Input Resistance
Input Voltage for 1 dB Compression
Maximum Safe DC Input Voltage
Maximum Safe RF Input Voltage
Frequency Range
a. When used with the active probe.

Inspection

Make sure the tip is not bent or discolored. Periodically inspect the barrel of the probe receptacle,
making sure it is clean and free of grit. Clean the receptacle with clean compressed air.
When cascading 10:1 dividers, periodically inspect and clean the exterior of the metal sleeve.
Chapter 2
10:1
Typically <0.7 pF
1 M
a
3.0 Volts Peak
200 Volts
a
15 Volts Peak
Same as the active probe (300 kHz to 3 GHz)