Precaution
On simultaneous use with an
electrosurgical knife, etc.
If this unit is used together with an
electrosurgical knife, etc., the picture may be
disturbed, warped or otherwise abnormal as a
result of strong radio waves or voltages from the
device. This is not a malfunction.
When you use this unit simultaneously with a
device from which strong radio waves or
voltages are emitted, confirm the effect of this
before using such devices, and install this unit in
a way that minimizes the effect of radio wave
interference.
Recommendation to Use more than One
Unit
As problems can occasionally occur for the
monitor, when the monitor is used for safety
control of personnel, assets or stable picture, or
for emergencies, we strongly recommend you
use more than one unit or prepare a spare unit.
On Cleaning
Before cleaning
Be sure to disconnect the AC power cord from
the AC outlet.
On cleaning the monitor
A material that withstands disinfection is used
for the front protection plate of the medical use
LCD monitor. The protection plate surface is
specially treated to reduce reflection of light.
When solvents such as benzene or thinner, or
acid, alkaline or abrasive detergent, or chemical
cleaning cloth are used for the protection plate
surface/monitor surface, the performance of the
monitor may be impaired or the finish of the
surface may be damaged. Take care with respect
to the following:
Clean the protection plate surface/monitor
surface with a 50 to 70 v/v% concentration of
isopropyl alcohol or a 76.9 to 81.4 v/v%
concentration of ethanol using a swab method.
Wipe the protection plate surface gently (wipe
using less than 1 N force).
Stubborn stains may be removed with a soft
cloth such as a cleaning cloth lightly dampened
with mild detergent solution using a swab
method and then clean using the above
chemical solution.
Never use solvents such as benzene or thinner,
or acid, alkaline or abrasive detergent, or
chemical cleaning cloth for cleaning or
disinfection, as they will damage the
protection plate surface/monitor surface.
Do not use unnecessary force to rub the
protection plate surface/monitor surface with a
stained cloth. The protection plate surface/
monitor surface may be scratched.
Do not keep the protection plate surface/
monitor surface in contact with a rubber or
vinyl resin product for a long period of time.
The finish of the surface may deteriorate or the
coating may come off.
To prolong the life of the unit
Turn off the power to preserve the performance
when not in use for a prolonged time.
Cautions for RESPONSIBLE ORGANIZATION
when connecting this equipment to IT-
NETWORK
– connection of the PEMS to an IT-NETWORK that
includes other equipment could result in
previously unidentified RISKS to PATIENTS,
OPERATORS or third parties;
– the RESPONSIBLE ORGANIZATION should
identify, analyze, evaluate and control these
RISKS;
– subsequent changes to the IT-NETWORK could
introduce new RISKS and require additional
analysis; and
– changes to the IT-NETWORK include:
changes in the IT-NETWORK configuration;
connection of additional items to the IT-
NETWORK;
disconnecting items from the IT-NETWORK;
update of equipment connected to the IT-
NETWORK; and
upgrade of equipment connected to the IT-
NETWORK.
13
GB