Removal/ Installation, Replacement and Adjustment
Advanced Valve System (AVS)
Before removing Bubble inject tubing turn off incoming Argon gas supply.
NOTE
If instrument is to be operated with Bubble inject supply disconnected, seal end of tubing with
joiner and blanking plug.
4 Remove Argon Bubble inject line by pressing black ring and pulling tubing in the direction of
the arrow.
5 Disconnect RJ45 connector.
6 Disconnect 2 x 24Volt supply lines by pushing tab and withdrawing connector.
7 Reassembly is the reverse of disassembly.