Call Qualification
Customer Help Guide
8 Ask Customer to light a plasma and check for conditional parameters not in the Red zone. If
in the Red zone ask the customer to adjust the relevant input pressure etc. Go to the
following sections to assist in diagnosis.
•
•
If no gauges in the Red zone progress to next steps.
9 For Software versions 7.3 and higher, Run Tests on "Tests" Tab. (For S/W versions below 7.3
only the "Instrument Performance" Test will be available).
Functional issues are outlined in red above. If any of these tests are failing this suggests an
instrument fault. Go to the following section for diagnosis.
•
10 Go to step 17 otherwise continue to next step.
Agilent 5110/5100 ICP-OES Service Manual
Dashboard descriptions
Instrument Specifications
Instrument Test Descriptions