Download Print this page

Emerson Rosemount 2130 Quick Start Manual page 25

Level switch
Hide thumbs Also See for Rosemount 2130:

Advertisement

May 2022
3.5.7
Fault condition detection (self-check mode only)
When a fault condition is detected in the self-check operating mode, the
heartbeat LED flashes once every half a second and every third flash is
missed. The output from the level switch will then be as
Table 3-15: Fault Condition Detection (Self-check Mode Only)
Direct load
1
2
3
< 4 mA
I L
R
N
L
0 V
+ V
(=Fault)
DPCO relay
NC C NO
NC C NO
(=Fault)
Fault and alarm (2 x SPCO) relays
Alarm relay
NC C NO
(=No alarm)
Quick Start Guide
1
Fuse
2A(T)
DPST
PLC
2
3
4
< 100 μA
I L
-
+
I/P
PLC
(=Fault)
NAMUR
1
2
< 1.0 mA
(-)
(+)
(=Fault)
Fault relay
NC C NO
(=Fault)
Quick Start Guide
Table
3-15.
PNP dc
1
2
3
4
R
I
L
Fuse
1A(T)
< 100 μA
(+)
(-)
(=Fault)
8/16 mA
1
2
3
< 3.7 mA
(-)
(+)
(=Fault)
= Load on
= Load off
25

Advertisement