Maico MA 28 Operation Manual page 51

Hide thumbs Also See for MA 28:
Table of Contents

Advertisement

Guidance and manufacturer's declaration - electromagnetic emissions
The MA 28 is intended for use in the electromagnetic environment specified below. The customer or the user of the MA 28 should
assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
portable and mobile RF communications equipment and the MA 28.
The MA 28 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the MA 28 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the MA 28 as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum
output power of
transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MA 28 is intended for use in the electromagnetic environment specified below. The customer or the user of the MA 28 should
assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Electrostatic
+8 kV contact
Discharge (ESD)
+15 kV air
IEC 61000-4-2
Electrical fast
+2 kV for power supply lines
transient/burst
+1 kV for input/output lines
IEC61000-4-4
Surge
+1 kV differential mode
IEC 61000-4-5
+2 kV common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
Voltage dips, short
40% UT (60% dip in UT)
interruptions and
for 5 cycles
voltage variations on
power supply lines
70% UT (30% dip in UT)
for 25 cycles
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
8506915 Rev. 5
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.17 √ ��
0.12
0.37
1.17
3.70
11.70
Compliance
+8 kV contact
+15 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT
for 5 sec
3 A/m
Operation Manual MA 28
Electromagnetic environment - guidance
The MA 28 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The MA 28 is suitable for use in all commercial, industrial, business,
and residential environments.
[m]
80 MHz to 800 MHz
d = 1.17 √ ��
0.12
0.37
1.17
3.70
11.70
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be greater than 30%.
Mains power quality should be that of a
typical commercial or residential
environment.
Mains power quality should be that of a
typical commercial or residential
environment.
Mains power quality should be that of a
typical commercial or residential
environment. If the user of the MA 28
requires continued operation during
power mains interruptions, it is
recommended that the MA 28 be
powered from an uninterruptable power
supply or its battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
residential environment.
50
800 MHz to 2.7 GHz
d = 2.23 √ ��
0.23
0.74
2.33
7.37
23.30
05/05/2021

Advertisement

Table of Contents
loading

Table of Contents