12.2 Programming density values

Before you begin
About this task
Steps
User's manual
NOTE
Make sure that no exposure region is selected, i.e. no preprogrammed
setting indicator light is on.
By changing the density value all the preprogrammed values can be
changed. This can be used for example when the cone is being changed.
1. Select the large or small animal mode of which the default exposure
values is to be programmed.
The indicator light of the selected mode will come on. The current
exposure values are shown on the time, kV and mA displays.
2. Press and hold down the SELECT key (about 4 seconds) until you have
heard a signal tone to enter the programming mode.
The ready light will start to flash. The time (s) display will start to flash
and the default exposure values appear on the displays.
The imaging mode can be changed by pressing the MODE key briefly.
3. Press the SELECT key briefly three times.
The density value can now be changed with the parameter adjustment
keys.
12 Programming exposure values