12 Programming exposure values
12.1 Programming default exposure values
Before you begin
About this task
Steps
User's manual
NOTE
The exposure values are programmed corresponding the density value 0
(factory preset value). The exposure time values are automatically scaled
according to the density value.
NOTE
Make sure that no exposure region is selected, i.e. no preprogrammed
setting indicator light is on.
The default exposure values can be programmed for both the large or small
animal mode.
NOTE
The exposure parameters - time, kV and mA - are programmed
corresponding to the density value 0. The time value will be automatically
changed according to the selected density value in the film-based imaging
mode, in the digital imaging mode and in the imaging plate mode when you
exit the programming mode.
For recommended default exposure values, see sections:
•
"Recommended exposure values for cats or smaller animals" on page 17
•
"Recommended exposure values for larger animals" on page 18
1. Select the large or small animal mode of which the default exposure
values is to be programmed.
The indicator light of the selected mode will come on. The current
exposure values are shown on the time, kV and mA displays.
12 Programming exposure values
Planmeca ProX
23