3.2 Other Components

3.3 Preparing for
Operation
Thermo Fisher Scientific
The following accessories are ordered separately:
Glass Condensation Flasks. GCF400 and GF4000
Order the wide-mouth GCF400 flask (4-liter Glass Condensation
Flask) for all standalone concentrators except for the SC250EXP.
Order the GF4000 for use only with the SC250EXP.
The glass flasks are quickly exchanged between runs and are easy to clean.
Having several available permits uninterrupted refrigerated trap operation
while the used vessel is being defrosted, cleaned and dried for next time.
®
CryoCool
Heat Transfer Fluid. A permanent, safe, efficient, and
economical alternative to methanol or ethanol. CryoCool does not
evaporate, has very low water absorption, is odorless, and non-toxic. Order
SCC1 for a 1-liter bottle; order SCC5 for a 5-liter bottle.
In addition, Thermo Scientific offers a complete line of other
components required for drying, including vacuum tubing kits,
SCT120 Chem traps and replacement Flask Covers for the GCF400.
Switch the trap OFF. Connect the power cord to the receptacle on the right
side of the instrument. Plug the trap into an appropriate wall outlet.
Before connecting the Refrigerated Vapor Trap to an
CAUTION!
outlet, check voltage, frequency, and amperage to be sure they match
the power requirements indicated on the rear of each instrument. (115
VAC/60 Hz, 13 A, 230 VAC/50 Hz, 7 A.) If there are any questions,
please consult an electrician.
As a safety feature, units are equipped with a three-prong grounded plug
that fits a grounding-type power outlet. Consult an electrician to replace
outlet if necessary. Do not defeat this safety feature by modifying the plug.
These units are "FOR INDOOR USE ONLY". Avoid operating in areas of
excessive humidity or extremes of temperature.
Pour CryoCool Heat Transfer Fluid (500 ml) into the stainless steel trap
chamber. Before each use, be sure the chamber is filled to the correct level. If
necessary. add more CryoCool.
Cryocool can now be drained from the RVT5105 system to remove
unwanted ice or other contaminates after the system has been fully
defrosted. Refer to Section 5 on page 10 for troubleshooting and
maintenance details.