CAL
The appliance has been calibrated at the factory. Please only calibrate, in
case the measuring results are wrong or the letters 'CAL' are blinking on
the display. A benchmark of 1.219 m above the base unit is important for
the gauging. Ideally, you should use the feeler rod for calibration.
Calibrating the ZipLevel:
1) Put the basic unit on its back onto a level surface and switch the
measuring module on.
2) Lay the measuring module on its base or back next to the basic unit.
Press the CAL-key for 2 seconds.
3) When (0) is blinking on the display, shortly press CAL.
4) When (1.219) is blinking on the display, raise the measuring module
to a height of 1.219 m above the basic unit and press CAL again. The
measuring module has to be held stable during the process. The feeler
rod or a tape measure mark at a wall should be used as the calibrating
standard.
5) When the display stops blinking and shows 1.219 m the appliance is
recalibrated.
!
You can leave the calibration function at any time without
calibrating! For that you press the CAL-key for 2 seconds.
14