10 Technical Data
Miscellaneous
Operating temperature range
Material
Mass
Recommended electronics
10.1.2
Limit Switch Specifications
Type
Supply voltage
Signal output
Signal logic
10.1.3
Reference Switch Specifications
Type
Supply voltage
Signal output
Signal logic
10.1.4
Maximum Ratings
M-687.UN XY stages are designed for the following operating data:
Maximum operating voltage
200 V
or 71 V
pp
38
Version: 1.3.0
Magnetic (Hall effect) sensor
+5 V / GND, supply via the motor socket
TTL level
Active high. When the limit switch is passed, the signal level
changes:
− Normal motor operation: low (0 V)
− Limit switch reached: high (+5 V)
Optical sensor
+5 V / GND, supply via the motor socket
TTL level
Direction sensing by means of different signal levels on the left-
and right-hand side of the reference switch: The signal level
changes from 0 to +5 V when the reference switch is passed.
Operating frequency
152 to 165 kHz
eff
MP100E
M-687.UN
XY Stage for Inverted Nikon
Microscopes
20 to 40
Aluminum (black anodized)
4.2
C-867.2U2
Unit
Tolerance
°C
kg
±5 %
Maximum power
consumption
60 W
M-687.UN XY Stage