5 Installation
Figure 7:
Requirements
 The adapter plate (p. 10) is mounted in the M-687.UN.
Tools and accessories
M-687.AP1 universal holder (p. 11)
Inserting the M-687.AP1 universal holder into the aperture
1. Align the universal holder so that the red dot on the holder points in the direction of
the clip spring of the aperture (refer to figure above).
2. On the side of the clip spring, place the short edge of the universal holder at a flat angle
in the middle of the aperture so that it makes contact on two sides.
3. Slowly slide the universal holder towards the clip spring until it reaches the edge and
the clip spring is compressed.
4. On the opposite side, press the universal holder slowly down until it snaps in
completely.
The universal holder is now clamped firmly.
Removing the M-687.AP1 universal holder from the aperture
 Slowly pull the universal holder on one side upwards until it releases.
5.6

Connecting the M-687.UN to the Controller

Requirements
 You have read and understood the general notes on installing (p. 15).
 You have read and understood the user manual for the controller (p. 3).
 The controller is switched off.
 You have connected the M-687.UN to the protective earth conductor (p. 16).
Tools and accessories
Two U-600.AMD adapter cables (p. 10)
22
Version: 1.3.0
M-687.UN: Clip spring in the aperture of the adapter plate
MP100E
M-687.UN XY Stage