7 Maintenance
7
Maintenance
In this Chapter
General Notes on Maintenance ................................................................................................... 31
Doing a Maintenance Run............................................................................................................ 31
Cleaning the M-687.UN ............................................................................................................... 32
7.1
General Notes on Maintenance
NOTICE
Damage due to improper maintenance!
Improper maintenance can result in the failure of the M-687.UN.
Loosen screws only when instructed in this manual.
Make sure that the piezo motors of the M-687.UN do not come into contact with
lubricants.
7.2
Doing a Maintenance Run
Depending on the operating conditions and the period of use of the M-687.UN, the following
maintenance measures are required:
Maintenance run
The maintenance run is done to distribute the existing lubricant on the guides of the M-687.UN.
To evenly distribute the existing lubricant on the stage guides, do a maintenance run
across the entire travel range after 500 hours of operation, or at least after 1 year.
If you move the M-687.UN over a small travel range (<20 % of the entire travel range) in
continuous industrial operation, do a maintenance run across the entire travel range
every 5000 motion cycles.
Lubrication
Under laboratory conditions, the guides of the M-687.UN only need to be lubricated in
exceptional cases. For continuous industrial use, the lubrication intervals must be defined
individually.
Do not lubricate the guides of the M-687.UN without consulting our customer service
department (p. 35).
Ensure that the piezo motors of the M-687.UN do not come into contact with
lubricants.
M-687.UN XY Stage
MP100E
Version: 1.3.0
31