7.1.8
Connect the cable with crimp terminal to the terminal block.
Never perform wiring with moisture on the product. Moisture or
other foreign matter entering the case may cause malfunction.
Do not perform wiring when the power is turned on. Otherwise
damage to the product and electric shock may result.
Never loosen any screws other than those on the terminal block.
Be careful not to lose any parts when wiring.
Tighten the M3.5 screws of the terminal block with a tightening
torque of 0.8 to 1.0 N·m.
When using a bare crimp terminal, keep a necessary insulation
distance with an insulation tube or others to prevent a live part
exposure for possible electric shock or short circuit (see
"Figure 7-3").
Connect so that the polarity is correct.
Be careful not to short-circuit the terminals.
7.1.9
Tighten the terminal box cover firmly to seal the terminal box.
The protection rating of this product is guaranteed only when
the components are properly assembled.
Ensure that the O-ring is tightly fitted to the seal surface of the
inner case (see "Figure 7-2"). Also check that there are no
scratches, breaks or dirt on the O-ring.
Tighten the O-ring so that it is not twisted over the entire
circumference.
Be careful not to get the wires caught when tightening.
Be careful not to scratch the seat surface of the inner case
(terminal box cover).
CAUTION
Live part
Insulation distance
< Figure 7-3 >
CAUTION
Insulation tube
19
TY-EJ15-003A Rev.2