Calibration Set-Up; Calibration Equipment - QED BIOGAS 5000 Operating Manual

Gas analyzer
Table of Contents

Advertisement

5K-MNL-BIOGAS5000
9.4

Calibration set-up

Warning
The regulator supplied with the calibration kit has been configured to deliver a fixed flow.
As the regulator's flow is factory set, it only requires a few turns to open, no adjustment is
necessary.
Warning
9.5

Calibration equipment

The diagram below displays the regulator and tubing equipment for user calibration:
Page 48
Do NOT attach the gas supply to the gas analyzer before
putting the analyzer into the 'Gas Check' screen. Select
'Check Spans' from the 'Operation Settings' menu.
Exhaust port
When the gas analyzer is being calibrated, there are two
possible exits for the gas, via the usual manner out of the
exhaust (yellow) port of the analyzer or in cases of over-
pressurization the 1/16
located on the regulator.
It is recommended that both ports have exhaust tubing
attached.
The exhaust tubing must emerge in a well-ventilated area.
Ensure there are no leaks in the tubing and connections.
The calibration of the gas analyzer should be carried out in a
safe area with all necessary precautions taken when using
potentially dangerous, explosive or toxic gases.
BIOGAS 5000 gas analyzer
"