MP30 MODEL asics Draw Down Meter PATENT PENDING Instruction Manual Part No. 95180 5-31-07 P.O. Box 3726 Ann Arbor, MI 48106-3726 USA 1-800-624-2026 Fax (734) 995-1170 info@qedenv.com www.qedenv.com...
Background Drawdown Control Quick Guide MP30 Cable connection Drawdown Control Point Selection Drawdown Control Response with MicroPurge basics controllers General Information on MP30 Features and Usage Equipment Check Use in the Field- Static Water Level Measurement Cleaning the Meter General...
2. The MP30 includes light and sound indicators of reaching the drawdown limit (dry probe condition); these function independently of the MicroPurgebasics controllers, and may also be used in this way with controllers other than Micropurge basics models.
1. Cable connects to identical ports on MP30 and basics controllers, labeled "Control Port". 2. Connect cable to MP30 after reel has been turned to lower probe to desired drawdown control point, to avoid twisting cable while tape reel turns.
If the probe isno longer submerged and it takes more than 10 minutes for the water level to recover such that the probe is submerged again, the MP30 willshut off.The unit will need to be turned back on again. In some cases the drawdown point in the well may be lowered to compensate for slowwell recovery.
GENERAL INFORMATION General Information on MP30 Features and Usage Equipment Check 1. Check equipment and battery by pressing the "ON" button. The unit should sound if the speaker switch is not switched off. Also, the "LEVEL SHUT OFF" light should be on if the probe is dry and the "DRAWDOWN CONTROL"...
GENERAL INFORMATION Use in the Field- Static Water Level Measurement 1. Reel the tape down the well carefully, with the "DRAWDOWN CONTROL" switched "OFF", avoiding the edge of the casing. Hang the unit on the casing with the provided triangular wire bracket where possible and run the tape over the frame leg to avoid cuts and nicks.
Continuous sound when unit is turned on or removed from water. 1. Make sure MP30 has "Drawdown Control" switched to "OFF" position. 2. Make sure probe conductor tip is clean. 3. Check for excess moisture on the back of the electronic panel and wire connector area.
Power: Battery Life: 30-40 Hours 150 or 300 Feet (46 or 91 m) Tape Length: Operating Temperature: -40° to +185° F (-40° to +85° C) For additional assistance contact QED Service at: Phone: 1-800-624-2026 1-734-995-2547 Fax: 1-734-995-1 170 E-mail: service@qedenv .com...
Any failure of the products to conform to this warranty will be remedied by QED in the manner provided herein. This warranty shall be limited to the duration and the conditions set forth below. Warranty duration is calc- ulated from the original date of purchase.
Page 12
RESPONSIBILITY OF THE PURCHASER The original Purchaser's sole responsibility in the instance of a warranty claim shall be to notify QED of the defect, malfunction, or other manner in which the terms of this warranty are believed to be violated. You may secure performance of obligations hereunder by contacting the Customer Service Department of QED and: 1.